Sunday, September 18, 2011

Magnehelic Pressure Gauge Series 2000



Mounting. A single case size is used for most ranges of Magnehelic gages. They can be flush or surface mounted with standard hardware supplied. With the optional A-610 Pipe Mounting Kit they may be conveniently installed on horizontal or vertical 1 1/4" - 2" pipe. Although calibrated for vertical position, many ranges above 1 inch may be used at any angle by simply re-zeroing.

However, for maximum accuracy, they must be calibrated in the same position in which they are used. These characteristics make Magnehelic gages ideal for both stationary and portable applications. A 4 9/16" hole is required for flush panel mounting. Complete mounting and connection fittings plus instructions are furnished with each instrument.

PHYSICAL DATA
  • Ambient temperature range: 20° to 140° F.*
  • Rated total pressure: -20" Hg. to 15 psig.+
  • Overpressure: Relief plug opens at approximately 25 psig
  • Connections: 1/8"' NPT female high and low pressure taps duplicated - one pair side and one pair back.
  • Housing: Die cast aluminum. Case and aluminum parts Iridite-dipped to withstand 168 hour salt spray test. Exterior finish is baked dark gray hammerloid.
  • Accuracy: Plus or minus 2% of full scale (3% on - 0 and 4% on - 00 ranges), throughout range at 70°F.
  • Standard accessories: Two 1/8" NPT plugs for duplicate pressure taps, two 1/8" pipe thread to rubber tubing adapter and three flush mounting adapters with screws. (Mounting and snap ring retainer substituted for 3 adapters in MP & HP gage accessories.)
  • Weight: 1 lb. 2 OZ.
  • * Low temperature models available as special option.
  • + For applications with high cycle rate within gage total pressure rating next higher rating is recommended. See Medium and High pressure options lower left.
More details will be written later.



Related Posts with Thumbnails